TY - JOUR
T1 - Note
T2 - Thermal analysis of the long line source electron gun
AU - Iqbal, M.
AU - Wasy, A.
AU - Lodhi, M. A.K.
PY - 2013/5
Y1 - 2013/5
N2 - We performed thermal analysis for our previously reported [M. Iqbal, K. Masood, M. Rafiq, M. A. Chaudhry, and F. Aleem, Rev. Sci. Instrum. 74, 4616 (2003)10.1063/1.1614852], long linear filament electron gun assembly using ANSYS software. The source was set under a thermal load of 3000 °C, to evaluate temperature distribution, thermal strain, and heat flux at various components of the gun. We calculated the maximum heat flux (9.0 W/mm2) that produced a thermal strain of 0.05 at the focusing electrodes. However, the minimum value of the heat flux (0.3 W/mm2) was at the anode electrodes which correspond to a negligible thermal strain. The gun was validated experimentally showing a uniform cross section of the beam at the molybdenum work plate comparable to the size of the filament. Our experimental and theoretical results are in agreement. The gun had been in continuous operation for several hours at high temperatures without any thermal run-out.
AB - We performed thermal analysis for our previously reported [M. Iqbal, K. Masood, M. Rafiq, M. A. Chaudhry, and F. Aleem, Rev. Sci. Instrum. 74, 4616 (2003)10.1063/1.1614852], long linear filament electron gun assembly using ANSYS software. The source was set under a thermal load of 3000 °C, to evaluate temperature distribution, thermal strain, and heat flux at various components of the gun. We calculated the maximum heat flux (9.0 W/mm2) that produced a thermal strain of 0.05 at the focusing electrodes. However, the minimum value of the heat flux (0.3 W/mm2) was at the anode electrodes which correspond to a negligible thermal strain. The gun was validated experimentally showing a uniform cross section of the beam at the molybdenum work plate comparable to the size of the filament. Our experimental and theoretical results are in agreement. The gun had been in continuous operation for several hours at high temperatures without any thermal run-out.
UR - http://www.scopus.com/inward/record.url?scp=84878920481&partnerID=8YFLogxK
U2 - 10.1063/1.4808331
DO - 10.1063/1.4808331
M3 - Article
C2 - 23742611
AN - SCOPUS:84878920481
SN - 0034-6748
VL - 84
JO - Review of Scientific Instruments
JF - Review of Scientific Instruments
IS - 5
M1 - 056113
ER -