Analysis of iTRAQ data using Mascot and Peaks quantification algorithms

Carla M.R. Lacerda, Lei Xin, Iain Rogers, Kenneth F. Reardon

Research output: Contribution to journalArticlepeer-review

18 Scopus citations

Abstract

The field of proteomics has been developing rapidly toward quantification of proteins. Despite the variety of experimental techniques available for peptide and protein labelling, there are few commercially available analytical tools with the ability to interpret data from any mass spectrometer. In this study, we compare two software packages, Mascot and Peaks, for the analysis of iTRAQ data from ESI-Q/TOF mass spectrometry. In the case of a six-protein mixture combined in a known proportion, the output of the Peaks algorithm deviated from the correct result by 14% on average, while the error of the Mascot quantification was nearly 200%. When the software were used to analyse iTRAQ data from a complex protein sample, the quantification results agreed within 205 for only 265 of the quantified proteins, showing significant differences in the two quantification algorithms. This comparison and analysis revealed major intricacies in peptide and protein quantification that must be taken into consideration for software development.

Original languageEnglish
Pages (from-to)119-126
Number of pages8
JournalBriefings in Functional Genomics and Proteomics
Volume7
Issue number2
DOIs
StatePublished - Mar 2008

Keywords

  • Mascot
  • Peaks
  • Quantitative proteomics
  • iTRAQ

Fingerprint

Dive into the research topics of 'Analysis of iTRAQ data using Mascot and Peaks quantification algorithms'. Together they form a unique fingerprint.

Cite this